The CD24AD99Q-4, CD24AD99Q-6, and CD24AD99Q devices are a family of four-, six-, and eight-channel, lownoise, 24-bit, simultaneous-sampling delta-sigma (ΔΣ) analog-to-digital converters (ADCs) with a built-in programmable gain amplifier (PGA), internal reference, and an onboard oscillator. The CD24AD99Q-x incorporates all commonly-required features for extracranial electroencephalogram (EEG) and electrocardiography (ECG) applications. With its high levels of integration and exceptional performance, the CD24AD99Q-x enables the creation of scalable medical instrumentation systems at significantly reduced size, power, and overall cost. The CD24AD99Q-x has a flexible input multiplexer per channel that can be independently connected to the internally-generated signals for test, temperature, and lead-off detection. Additionally, any configuration of input channels can be selected for derivation of the patient bias output signal. Optional SRB pins are available to route a common signal to multiple inputs for a referential montage configuration. The CD24AD99Q-x operates at data rates from 250 SPS to 16 kSPS. Lead-off detection can be implemented internal to the device using an excitation current sink or source. Multiple CD24AD99Q-4, CD24AD99Q-6, or CD24AD99Q devices can be cascaded in high channel count systems in a daisy-chain configuration.
The CD24AD99Q-x is offered in a TQFP-64 package specified from –40°C to +85°C.
- Resolution (Bits):24
- Channels:8
- Interface type:SPI
- Power Consumption:39 mW/span>
- Eight Low-Noise PGAs and Eight High-Resolution ADCs
- Low Power: 0.75 mW/channel
- Input-Referred Noise: 4 μVPP (150 Hz BW, G = 6)
- Input Bias Current: 200 pA Data Rate: 250 SPS to 32 kSPS
- CMRR: –115 dB
- Programmable Gain: 1, 2, 4, 6, 8, 12, or 24
- Supports systems meeting AAMI EC11, EC13, IEC60601-1, IEC60601-2-27, and IEC60601-2-51 Standards
-
Unipolar or Bipolar Supplies:
- AVDD = 2.7 V to 5.25 V
- DVDD = 1.65 V to 3.6 V - Built-In Right Leg Drive Amplifier, Lead-Off Detection, Wilson Center Terminal, Pace Detection, Test Signals
- Integrated Respiration Impedance Measurement
- Digital Pace Detection Capability
- Built-In Oscillator and Reference
- SPI™-Compatible Serial Interface
-
Electroencephalogram (EEG) Study
-
Fetal Electrocardiography (ECG)
-
Sleep Study Monitor
-
Bispectral Index (BIS)
-
Evoked Audio Potential (EAP)